Siemens AG Healthcare Sector
11/12
Guidance and manufacturer’s declaration EMC
8
A MR system is classified as a large and permanently-installed equipment. The exemption
according to the requirement 36.202.3 b) 9. of IEC 60601-1-2 Edition 2.1 (2004-11) and
6.2.3.2i of IEC 60601-1-2 Edition 3 (2007-03) has been used and the equipment was not
tested for radiated RF immunity over the entire frequency range 80 MHz to 2.5 GHz.
!
The MR system has been tested for radiated RF immunity only at selected frequencies.
Tested ISM frequencies
80 MHz–2.5 GHz
Modulation characteristic
1 kHz, 80% AM
Test level
3 V/m
Signal generator with amplifier
9 kHz to 3.2 GHz
Содержание MAGNETOM Skyra
Страница 2: ......
Страница 4: ......
Страница 8: ...Siemens AG Healthcare Sector 11 12 Introduction Page 4 of 4 ...
Страница 12: ...0 0 Safety instructions A 2 0 0 MAGNETOM Family Safety Information0 0 ...
Страница 37: ...A 1 ...
Страница 39: ...MR MAGNETOM System Owner Manual MR compatibility data sheet Skyra ...
Страница 62: ...Siemens AG Healthcare Sector 11 12 MR compatibility data sheet 24 ...
Страница 63: ...MR MAGNETOM System Owner Manual Guidance and manufacturer s declaration EMC Skyra ...
Страница 73: ...MR MAGNETOM System Owner Manual Technical data Skyra ...
Страница 102: ...Siemens AG Healthcare Sector 11 12 Technical data 30 ...
Страница 103: ...MR MAGNETOM System Owner Manual Location of labels Skyra ...
Страница 105: ...3 Siemens AG Healthcare Sector 11 12 Location of labels Patent label ...
Страница 107: ...5 Siemens AG Healthcare Sector 11 12 Location of labels Safety certificate label U S A only ...
Страница 108: ...Siemens AG Healthcare Sector 11 12 Location of labels 6 ...
Страница 123: ...MR MAGNETOM System Owner Manual Disposal Skyra ...