7.3.5.3
Teaching in the zero point
The current distance is taught in as a new zero point (reference point) or reset to the
output value.
7.3.5.4
Setting the sampling duration
The following settings are possible:
•
12.5
μ
s
•
14.3
μ
s
•
16.7
μ
s
•
20
μ
s
•
25
μ
s
•
50
μ
s (default value)
•
100
μ
s
•
500
μ
s
•
1,000
μ
s
•
AUTO (selects the sampling rate for the quickest possible measurement depending
on the measuring object)
NOTE
If the sampling duration is set to a time period of
≤
20
μ
s , then the measuring range is
reduced.
■
Select the required range (near, medium, far) via the configuration interface.
For a sampling duration of
≤
20
μ
s
Table 12: Measuring range at sampling duration
≤
20
μ
s (OD5000-C15xxx and OD5000-C30xxx)
OD5000-
C15T01
OD5000-
C15W01
OD5000-
C30T05
OD5000-
C30W05
Near
14.0 mm to 14.6 mm
25.0 mm to 28.1 mm
22.5 mm to 24.0 mm (for specular
reflection)
Medium (starting
value)
14.4 mm to 15.4 mm
27.8 mm to 31.9 mm
22.8 mm to 27.9 mm (for specular
reflection)
Dist.
15.3 mm to 16.0 mm
31.1 mm to 35.0 mm
26.7 mm to 28.5 mm (for specular
reflection)
Table 13: Measuring range at sampling duration
≤
20
μ
s (OD5000-C85xxx and OD5000-
C150xxx)
OD5000-
C85T20
OD5000-
C85W20
OD5000-
C150T40
OD5000-C
150W40
Near
65.0 mm to 77.7 mm
71.5 mm to 74.3 mm (for specular
reflection)
110.0 mm to 134.4 mm
Medium (starting
value)
73.5 mm to 90.8 mm
70.6 mm to 86.9 mm (for specular
reflection)
124.8 mm to 166.3 mm
Dist.
84.8 mm to 105.0 mm
81.0 mm to 91.5 mm (for specular
reflection)
150.2 mm to 190.0 mm
An upper and lower limit must be set for the
Auto
setting. The actual sampling duration
is automatically adjusted within these limits.
OPERATION
7
8021391/17Y6/2020-05-11 | SICK
O P E R A T I N G I N S T R U C T I O N S | OD5000
37