5.3.1
Function of automatic adjustments/validations
8 adjustment or validation procedures that can run fully automatically can be pro‐
grammed. Each procedure uses one of the test gases configured in the Test Gas Table
(
see "Configuring test gases (Test Gas Table)", page 39
).
Programming defines which measuring component is to be adjusted or validated with
the procedure and which adjustment or validation function is to be used. Several com‐
ponents can be adjusted in one function.
5.3.2
Start options
•
Manual start:
Use menu
Operator commands
).
•
Automatic start (in regular intervals):
Use a long-term timer in the start conditions (see “Technical Information BCU”).
•
Remote-controlled start:
Control the value of a Boolean variable (VBVi) with a digital input (see “Technical
Information BCU”).
•
Programmed start conditions:
Calculate the value of a Boolean variable (BVIi) with a formula (see “Technical
Information BCU”).
For representation of the adjustment and validation results (see “Technical Information
BCU”).
5.3.3
Programming automatic adjustments/validations
1.
Menu: Call up
BCU
/
Parameter
/
Adjustment/Validation
.
Figure 24: Menu Adjustment / Validation – Table (example)
1
Index number of the adjustment or validation procedure
2
Programmed name
3
= automatic starts activated
4
Start time of next automatic function
5
Start date of next automatic function
6
Component X that will be addressed by the respective function
7
Number of the next function to start automatically (only when procedures are
chained)
2.
Deactivate
Live view
.
3.
Select the desired Table rows.
4.
Select
Edit
.
ADJUSTMENT FUNCTIONS
5
8017324/15A2/V6-0/2019-10 | SICK
T E C H N I C A L I N F O R M A T I O N | MARSIC200
45