O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER C200
Subject to change without notice
4
START-UP AND PARAMETER SETTINGS
4.4
Setting standard parameters
4.4.1
Assigning the MCU to the sender/receiver unit
The MCU must be set to the sender/receiver unit to be connected. A malfunction is
reported in case of a mismatch. Assignment must be made after installation when the
setting is not possible at the factory (e.g., when several devices are delivered at the same
time or when the MCU is swapped later). The following steps are then necessary:
▸
Connect the measuring system to the SOPAS ET program.
▸
Enter the Level 1 password (
see “Password and operating levels”, page 84
▸
Set the sender/receiver unit to “Maintenance”: Click “Maintenance sensor”.
Fig. 44: SOPAS ET menu: MCU/Maintenance/Maintenance
▸
Change to the “Configuration / Application selection” directory (
MCU/Configuration/Application selection”, page 68
▸
The basic type of the sender/receiver unit connected is displayed in the “Connected
variant” window (field “Application selection”). Click “Save selection” to assign to the
MCU
.
Fig. 45: SOPAS ET menu: MCU/Configuration/Application selection
The sender/receiver unit must be connected to the MCU.