99
8011952/YWL2/3-0/2016-08| SICK
O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER C200
Subject to change without notice
MAINTENANCE
5
▸
Trigger the function check (select the “Adjustment / Function Check - Manual”
subdirectory in the “MCU“ device file and click “Start Manual Function Check”;
“SOPAS ET menu: MCU/Adjustment/Function Check - Manual”, page 95
▸
Check the contamination value (
see “Maintenance on the sender/receiver unit and
reflector/scattered light receiver”, page 94
see “SOPAS ET menu: DH C200/Diagnosis/
▸
If the measured values for contamination, zero point and span are within the allowed
ranges, save them to the device by clicking the “Refresh” button (“Check values” field); if
not, repeat cleaning and check the contamination value again by triggering a renewed
function check.
▸
Put the pivoted shutter cover back on, remove the cover from the assembly flange, re-
install and fasten the scattered light receiver optics.
▸
Swivel the enclosure back again and lock with the knurled screws.
▸
Move the pivoted shutter back into the measuring position. To do this, click the
“Measurement” button under “Pivoted shutter reflector” in the “Adjustment / Manual
adjustment / Motor control” directory (
see “SOPAS ET menu: DH C200/Adjustment/
Manual Adjustment/Motor control”, page 94
).
▸
Resume Measuring mode.