O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER C200
Subject to change without notice
START-UP AND PARAMETER SETTINGS
4
Fig. 34: Assembly on a dust-free path with zero tube
▸
Using the associated connection lines, connect the sender/receiver unit to the MCU and
the reflector/scattered light unit to the sender/receiver unit.
▸
Connect the MCU to the supply voltage.
▸
Start the SOPAS ET program and connect to the measuring system (
the device via USB line”, page 55
▸
Enter the Level 1 password (
see “Password and operating levels”, page 84
).
▸
Set the sender/receiver unit to “Maintenance”: Click “Maintenance sensor”.
Fig. 35: SOPAS ET menu: DH C200/Maintenance/Maintenance
▸
Clean the optical surfaces on sender/receiver unit and reflector (
the sender/receiver unit”, page 94
see “Maintenance on the reflector”, page 98
▸
Wait approx. 30 min before starting the following work (measuring system must have
reached operating temperature).
A
Ø76
X
X: Dimensions between flange and tube according to the
type of reflector/scattered light receiver (
scattered light receiver”, page 115
,
Sender/receiver unit
Reflector/scattered light receiver
Adjusting stand