72
8011952/YWL2/3-0/2016-08| SICK
O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER C200
Subject to change without notice
4
START-UP AND PARAMETER SETTINGS
Optional Ana-
log Output
Modules
Use first analog
output module
Inactive
Not permitted for DUSTHUNTER C200 (results in error, because AO
2 and AO 3 are available by default).
Active
Opens the fields to set parameters for AO 2 and AO 3 (standard on
DUSTHUNTER C200)
Analog Output
1 Parameter
Value on analog
output 1
Conc. a.c. (SI)
Dust concentration under
operating (actual) conditions
(based on scattered light intensity)
The selected measured
variables are output on the
analog output.
Conc. s.c. (SI)
Dust concentration under standard
conditions (based on scattered
light intensity)
Opacity
Extinction
Transmission
SI
Scattered light intensity
Rel. opacity
Relative opacity
Conc. a.c. (ext)
Dust concentration under
operating (actual) conditions
(based on extinction)
Conc. s.c. (ext)
Dust concentration under standard
conditions (based on extinction)
Live zero
Zero point
(0, 2 or 4 mA)
Select 2 or 4 mA to ensure being able to differentiate between
measured value and switched off device or interrupted current
loop.
Output
checkcycle
results on the
AO
Inactive
Control values (
) are not output on
the analog output.
Active
Control values are output on the analog output (the “Output
control values at AO” checkbox in the “Adjustment / Function
Check - Automatic” directory must be activated).
Write absolute
value
Inactive
Positive and negative measured values are differentiated.
Active
The amount of the measured value is output (not relevant for
DUSTHUNTER C200).
Analog Output
1 Scaling
Range low
Lower measuring
range limit
Physical value at live zero
Range high
Upper measuring
range limit
Physical value at 20 mA
Field
Parameter
Remark