START-UP AND PARAMETER SETTINGS
4
Group
Input window
Parameter
Remark
Device
identification
Mounting
location
Name of the measuring
location
Assignment of measuring system to the respective
measuring location
Flange - flange
Flange distance measured
on the duct
For registration (e.g. repeated reference setting on a
dust-free path)
Opt. measuring
distance
Active measuring path
Input required for calculation of relative opacity
Chimney
opening
Chimney diameter at top
end
Correction
factor
Value
Adaption of relative opacity to geometric dimensions of
channel
Concentration
calibration
coeffecients = f
(extinction)
cc2
Quadratic coefficient
Input of regression coefficients determined on
extinction basis for measurement of the dust
concentration in mg/m³ (
concentration measurement based on extinction”,
).
cc1
Linear coefficient
cc0
Absolute coefficient
Concentration
calibration
coeffecients = f
(Scattered light)
cc2
Quadratic coefficient
Input of regression coefficients determined on
scattered light basis for measurement of the dust
concentration in mg/m³ (
concentration measurement based on scattered light
).
cc1
Linear coefficient
cc0
Absolute coefficient
Primary
measuring
principle
Primary
measuring
principle
Transmission
Determination of light source (LED for 'Transmission‘,
laser for 'Scattered light‘) for determination of the check
values during the function check
Scattered light
RK_Gruppe_LED_
Span2
Span2
transmission
Activated
The second control value is
compared with the reference
value during the function check.
Activation and input
of reference value
possible only when
the “EPA Conformity”
is activated
Deactivated
The second control value is not
used.
Reference value Transmission value in %
Input of a second control value
for the function check (
)
Limit
contamination
and average
Limit
contamination
Value in %
Determination of a limit value, optionally 40% (default
value), 20%, 10%, 6%, 4%
When checkbox “EPA Conformity” is activated, limit
value 4% is firmly assigned.
Limit warning
75% of limit value
Automatic determination depending on the entered
limit value
Average active
Activated
When the checkbox is activated, an average is created
from the values measured in the determined average
interval.
Deactivated
Average Interval Interval time
1/2/3/4/5/6min
Selection of interval time (6 min for use according to
EPA standard)
Selection
Measure Value
Measured variable
Selection of measured variable whose values are to be
averaged
EPA Conformity
EPA-mode
active
Activated
For use according to EPA standard
Deactivated
No use according to EPA standard
After deactivation of the EPA conformity, limit contamination and limit warning are reset
to standard values (40% or 30%).