START-UP AND PARAMETER SETTINGS
4
4.4.4
Setting the analog outputs parameters
Select the “Configuration / IO Configuration / Output Parameters” directory to set the
analog outputs.
Fig. 47: SOPAS ET menu: MCU/Configuration/IO configuration/Output Parameters
●
Default values
see “Factory settings”, page 69
●
In order to output the dust concentration under standard conditions (“Conc. s.c.”
(Ext)), set the parameters for the analog outputs according to
.
Field
Parameter
Remark
Analog Out-
puts
- General
Configuration
Output Error
current
Yes
Error current is output.
No
Error current is not output.
Error Current
Value < Live Zero (LZ)
or > 20 mA
mA value to be output in “Malfunction” state (error case) (size
depends on connected evaluation system).
Current in
maintenance
User defined value
A value to be defined is output during “Maintenance”
Value measured last
The value measured last is output during “Maintenance”
Measured value
output
The current measured value is output during “Maintenance”.
Maintenance
current
Whenever possible,
value
≠
LZ
mA value to be output in “Maintenance” state