61
8011952/YWL2/3-0/2016-08| SICK
O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER C200
Subject to change without notice
START-UP AND PARAMETER SETTINGS
4
▸
Insert the screw driver in the focussing opening and set the adjustment screw so that
the scale pointer in the control window points to the distance between the optical
surfaces of the sender/receiver unit and the reflector/scattered light receiver
(= distance of the front panels).
Active measuring path 0.5 ... 3 m:distance = dimension A + 343 mm
Active measuring path 2.5 ... 8 m:distance = dimension A + 348 mm
(
see “Assembly on a dust-free path with adjusting stands”, page 58
)
Fig. 38: Focussing the sender light beam
▸
Screw the cover for the focussing opening on again, swivel the electronics unit back and
lock with the knurled screws.
▸
Click the “Mechanical centring” button (step 1‘) in directory “Adjustment / Manual
Adjustment / Transmission set reference” (
see “SOPAS ET menu: DH C200/Adjustment/
Manual Adjustment/Transmission set reference”, page 60
).
▸
Align the optical axes of the sender unit and the scattered light receiver to each other.
Align the sender/receiver unit so that the sender spot lies on the middle of the opening
for the reflector (
see “Reflector/scattered light receiver for short measuring paths”,
). Align the reflector/scattered light receiver so that sender spot (1) can be seen
in the circular marking in the middle of the control window (2) on the back of the
enclosure (3).
The scale illumination lights when the measuring system is switched to “Mainte-
nance” or up to 10 min after device reboot.
4-quadrant display
Focussing opening
Scale
Control window
Reduce active measuring path
Electronics unit
Increase active measuring path