Designation
Syntax
Permanently switch the
heater off
1
--> Response OK
<STX>sMN switchHeaterOff<ETX>
<STX>sAN switchHeaterOff<ETX>
Heater automatic mode
--> Response OK
<STX>sMN switchHeaterAuto<ETX>
<STX>sAN switchHeaterAuto<ETX>
Writing preset values
--> Response OK
Value coded in hexa‐
decimals, unit in mm,
-1500 m ... 1500 m, cor‐
responds to FFE91CA0 ...
16E360, note: 32 bit!)
<STX>sWN preset <n*[0…F]><ETX>
<STX>sWA preset<ETX>
Read preset
--> Response OK
<STX>sRN preset<ETX>
<STX>sRA preset <n* [0...F]><ETX>
Activating preset
--> Response
Distance output value is
set to the preset value for
the current distance
<STX>sMN activatePreset<ETX>
<STX>sAN activatePreset<ETX>
Setting output to 0
--> Response OK
Distance value output is
set to 0 for the current dis‐
tance
<STX>sMN autoZero<ETX>
<STX>sAN autoZero<ETX>
Deleting preset value
--> Response OK
Distance output value cor‐
responds to the real meas‐
ured value, offset = 0
<STX>sMN resetPreset<ETX>
<STX>sAN resetPreset<ETX>
Restart
--> Response OK
Complete reboot
<STX>sMN mSCreboot<ETX>
<STX>sAN mSCreboot<ETX>
Read out interface version
--> Response
<STX>sRN interfaceVersion<ETX>
<STX>sRA interfaceVersion B 001.004.019<ETX>
Read out firmware creation
date
--> Response
<STX>sRN firmwareBuildTime<ETX>
<STX>sRA firmwareBuildTime 14 2016/04/19
16:12:23<ETX>
Read out firmware verifica‐
tion key
--> Response
<STX>sRN firmwareVerification<ETX>
<STX>sRA firmwareVerification 13
1106-0684-9500-0000<ETX>
Reset to factory settings
--> Response OK
<STX>sMN resetParamAndReboot<ETX>
<STX>sAN resetParamAndReboot <[0, 1]><ETX>
1
The heater can be activated via SOPAS ET, the RS-422 interface, the PROFIBUS DP interface or the
Ethernet interface. When the heater is switched on and the upper or lower temperature warning limit is
exceeded, the heater automatically switches to automated mode.
3.7.3.2.5
Examples of command sequences
Switching on continuous, permanent distance value output
Table 18: Example: Switching on continuous, permanent distance value output
To Dx1000
<STX>sMN SetAccessMode 4 81BE23AA<ETX>
From Dx1000
<STX>sAN SetAccessMode 1<ETX>
To Dx1000
<STX>sWN rs422PeriodicOutputContent 1<ETX>
PRODUCT DESCRIPTION
3
8024560/1A19/2021-06-30 | SICK
O P E R A T I N G I N S T R U C T I O N S | DT1000 PROFIBUS DP and DL1000 PROFIBUS DP
25