Operating instructions Distance measuring device DL100 Pro – RS-422
Operation at the measuring device
52
© SICK AG • Subject to change without notice • 8014753/ZVZ7/2018-03-22
8.5.13 Menu "MF2"
This menu and the associated submenus can be used to set parameters
for the multifunction output MF2.
You can get to the menu "MF2" via the menu path:
Main menu
→
→
Menu
→
→
RS422
→
→
more
→
→
MFx On
→
→
MF1
→
→
MF2
Push the
key for at least 2 seconds to get to the "Menu".
Push the
key so that the parameter "ActSta." is displayed.
Use the keys
and
to browse within the menu. Push the
key to
display the respective parameter value.
Requirements for the display
• Menu "more": Option "On"
• Menu "MFx On": Option "Enable"
Parameter
Description
ActSta
Select multifunction output level MF2.
Options
• ActLow: LOW-level at active output (opener/NC)
• ActHi: HIGH-level at active output (closer/NO)
Factory setting
• ActLow
Functn
Select function for the multifunction output. Depending on the
selection, the corresponding submenu is displayed.
Options
• Dist
• Srvice
• Speed
Factory setting
• Dist
Dist / Speed /
Srvice
Depending on the selection for the parameter "Functn",
the corresponding submenu is displayed. For parameter
description, see the respective table.
Count
Counts the switching events of the multifunction output.
Reset the counter by switching the distance measuring device
off and on again.
Table 18: Menu "MF2
"
Submenu "MF2 – Srvice"
This submenu corresponds to the submenu "Srvice" in the menu "MF1".
→
Submenu "MF2 – Dist"
This submenu corresponds to the submenu "Dist" in the menu "MF1".
→