Chapter
8
Operating Instructions
CLV 490 Bar Code Scanner
8-8
©
SICK AG · Division Auto Ident · Germany · All rights reserved
8 008 796/0000/25-06-2002
Troubleshooting
6. "error: writing parameters
CLV -> EXT,
internal parameters
loaded"
(Error type: 110, 111,
112)
Start option: S
TART
WITH
C
OPY
I
NTERN
–> E
XTERN
.
The CLV was
not
able to copy
the internal parameter set to
the external memory. It has
loaded the internal parameter
set to its RAM
Data transmission error.
Check whether the external para-
meter memory is connected cor-
rectly (free wiring: check wire
assignment).
Repeat the startup procedure. If
the error is still present, contact
the SICK Service department.
7.
"warning: additional
unknown parameter(s),
external parameters
loaded"
(Error type: 101, 114)
Start option: S
TART
WITH
E
XTER
-
NAL
P
ARAMETERS
.
The CLV has successfully
loaded the external parameter
set with tolerated errors
1)
to its
RAM.
Transfer ok but different para-
meter versions in the external
memory and the CLV.
Upload the parameter set to CLV-
Setup and check the values
1)
.
1) see footnote 1) in
Message
Meaning
Possible cause
Remedy
Table 8-3: Messages for problems accessing the external parameter memory