85
Example: Setting the terminal temperature for Channel 1.
Enter engineering mode (See the key operation on page
69). Use the
key to display "
", and press
the “ENT” key.
Use the
key to select "
", and press the “ENT”
key.
Select a channel to be set, and press the “ENT” key.
Measures and displays the current terminal temperature.
Press the “ENT” key, and the unit moves to temperature
input display.
Use the
key to change a numeral, and
key to
shift a digit. Then, press the “ENT” key. See Section 9.5
(3) Calibration of Reference Junction Compensation.
If calibration is performed correctly, select "
"
with the
key, and if calibration is performed
incorrectly, select "
" with the
key.
Then, press the “ENT” key.
7.3 Terminating the Engineering Mode
Example: The following describes how to save the setting data.
Use the
key to display "
", and press the “ENT”
key.
Use the
key to select "
", and press the
“ENT” key. The settings of the engineering mode have
been saved. Recording starts under the new settings.
[Note]
For terminal temperature calibration, select a channel from Ch1 to Ch3, and select another channel
from Ch4 to Ch6. Calibrate these two channels separately.
[Note]
When terminating Engineering mode, if "
" is selected and the power was turned off without
selecting "
", the settings will become invalid.
Selecting "
" invalidates the settings, and starts recording under the previous settings.
ENT
ENT
Display
Operation keys
Description
ENT
ENT
ENT
ENT
ENT
ENT
Display
Operation keys
Description
ENT
FEED