Basic Operations
Basic Operations
Switching between Displays
Switching between Minimum Number of
Displayed Digits (1d/10d Display)
It is possible to erase one digit for a minimum display (displays 10d).
1d Display
10d Display
Press [ 1d/10d] when in the weight measurement mode
The minimum number of displayed digits will alternate.
When taking a measurement that exceeds the capacity of the small weight range on a dual
range model, it cannot switch to 10d (0.001 g).
Switch measurement range (Only AP125WD-AD,
AP225WD-AD, AP125WD, AP225WD)
When the power supply and the switch are turned ON, the minimum display is set to the “Small weight
range” at 0.00001 g. To set the minimum display to the “Large weight range” at 0.0001 g, press [
1d/10d] in the weight display to switch the measurement range.
If a sample is measured that exceeds the capacity (AP125WD: 52g and AP225WD: 102g) of the small
weight range (while the small weight range is being used), the scale automatically changes to the
large weight range. In this situation, if the [
O/T] is pressed in large weight range with the tare
weight, this setting is calibrated for the large weight range. Even if the load on the pan is returned to
the capacity of the small weight range, the display does not return back to the small weight range.
Press the [
O/T] at this time to switch to small weight range.
(When using AP225WD)
Small weight range
Large weight range
Unit change
It is possible to switch between the pre-registered units.
Press [
UP] when in the weight
measurement mode
The pre-registered units will alternate every time this
is pressed.
Unit setting
Only [g] is registered as the default. When other
units are required, it is necessary to register them in
advance.
^
Unit Display after Rebooting
The unit that was used most recently will be displayed
when the power is switched off and the balance rebooted.
*
*
*