Calibration
Calibration
4.
Place a weight with the weight
displayed on the pan
(1) Make sure that nothing is on the pan.
(2) Place a weight with the weight displayed on the
pan.
OK
FUNC
g
d
5.
Remove the weight from the pan
Remove the weight from the pan.
The screen will return to the weight measurement
mode when adjustment is complete.
OK
FUNC
g
d
CAUTION
Instructions
Do not leave or move the balance if adjustment or tests have not been
completed or ended normally
The internal weights are not maintained, so moving the balance may result in malfunctions. Make sure that the power is
switched off and on once so that the balance can be started up normally (with the internal weights maintained).
Not applicable to a verified balance as a legal measuring instrument in the EU.
*