Attaching the Parts (Semi-micro models)
1.
Attach the adapter cable holder
Peel the protective sheet of adhesive off the adapter cable holder,
and stick it on the back of the balance as shown in the figure.
2.
Attach the shield plate
Align the hole in the center of the shield plate with the position of the
protrusion around the pan shaft and set it in place slowly and gently
from above.
3.
Put the pan supporter in place
Gently attach the pan supporter on the center axis of
the weighing chamber with hollow (two places) of it.
CAUTION
There is a risk of damage when placing the pan supporter, so make
sure to hold it by the depressions and place it gently.
4.
Put the pan on the pan supporters
Align the two pan notches with the left and right on
the balance main body.
5.
Set the protective cover
Make sure to attach the protective cover. Attaching the protective cover protects the panel sheet
from being worn out. (Be especially sure to attach the protective cover for the IR sensor equipped
AP W-AD Series in particular to prevent IR sensor malfunctions.)
(1) Peel the double-side tape from the protection cover (five locations).
(2) Place the protection cover over the balance’s operation keys and display.
(3) With the protection cover in place on the operation keys/display panel, press down firmly on
the double-side tape until it lodges in place.
CAUTION
If the protective cover rises up off the operation keys, it can result in malfunctions, so make sure to fix it in place securely
using double-sided tape. In addition, if the protective cover touchless sensor is dirty, it will decrease sensor sensitivity,
and could result in malfunctions, so wipe it clean or replace it so that it is not dirty.
CAUTION
Remove the pan, pan supporter, shield plate and inner draft shield plate without fail when repairing or moving the
balance. Failure to observe this may result in damage.
^
[Precautions when moving the balance or shipping the balance for repair] (P.157)
Using a verified balance as a legal measuring instrument in the EU:
Legal regulations require a verified balance be sealed. This control seal is a
self-destructive
adhesive label. This seal is irreparably damaged invalidating
the verification, if you attempt to remove
it. The balance must then be re-verified before it is used for legal measurements.