Calibration
Calibration
Performing Periodic inspections
It is recommended that repeatability inspections, corner load error inspections and instrumental error
inspections (performance inspections) are carried out periodically. Although these inspections may be
carried out independently, they can also be carried out in series, as explained below.
Precautions
• These performance inspections may be considered to be a yardstick for determining if the balance is operating normally.
• Carry out the performance inspections in a location free from sudden changes in room temperature.
• Place the weights inside the weighing chamber in advance to make sure the inspections are carried out without any
differences in temperature between the weighing chamber and the weights.
• Use elongated pincers to place and remove the weights, and do not put hands inside the weighing chamber.
1.
Call out the Periodic inspections
Press [
CAL] for three or more consecutive seconds.
2.
Start the Periodic inspections
Select [Periodic inspection], and then press [
OK].
3.
Perform the repeatability inspection
(1) Select [Repeatability Inspection], and then press
[
OK].
(2) Select [Start] to commence the repeatability
inspection.
(3) Place the designated weight on the pan and wait
for the display to change.
Timing for Display Change
The balance will automatically change the display
when the weight is placed or removed from the pan.
It takes the stable time for about 30 seconds for
automatic data output.
(The stable time differ from the model.)
The width of stability detection is set at ±1 count
(default setting when shipped from the factory).
(4) Remove the weight from the pan as the direction.
(5) Repeat Steps 3 and 4 in accordance with the on-
screen directions.
The repeat count is set at 6 (default setting when
shipped from the factory).
(6) Check the results of the inspection, which will be
displayed when the inspection is complete.
(7) Press [
OK].
The width of stability detection is set at ±1 count
(default setting when shipped from the factory).
When [Incompatible] is displayed...
Check to ascertain that the set environment is compatible and that there are no abnormalities
with the balance, and then run the inspection once more.
Judges whether or not the results of the inspection are within the permissible range of error for
the measurement values (maximum value - minimum value).
*
*