CAUTION
Instructions
Use the correct power supply and voltage with the AC adapter supplied.
Using the balance with an incorrect power supply or voltage may result in the outbreak of fire or malfunctions. Note
also that if the power supply or voltage is unstable or if the power supply capacity is insufficient, it will not be possible to
obtain satisfactory performance from the balance.
Warming Up
Warming-up is performed to ensure that the balance is as stable as possible during adjustment
and precise measurements. Leaving the balance with the power on in the weight measurement
status or in the stand-by mode (
P.34) for several hours will stabilize the internal
temperature of the balance. This procedure is known as warming-up.
Warming up requires the power to be left switched on for at least one hour.
The time required for warming up varies depending on the model.
Semi-micro models: At least 4 hours
Other models:
At least 1 hour
Log-in Function
^
The balance can be used with the Log-in Function.
The weight measurement screen will be displayed when the Log-in Function is not used.
3.
Date/Time Confirmation
Refer to the following for details on adjusting the clock:
4.
Carrying Out Simple Measurements
Press the key on the control panel to perform simple measurements.
[Zero Setting/Taring]
Place the measurement container on the pan and carry out taring (zero setting). Or,
press the key to return the display to zero.
After the [Weighing] zero display:
Place the sample to be weighed on the pan, and the value displayed when the
stability mark is shown is the result of the measurement.
*
*