USER MAINTENANCE (CONTINUED)
Continued on next page
Temperature Calibration (Continued)
5. Enter the Configuration menu.
a. Press the Program button to enter the Configuration menu.
b.
The Input option “InP” will be the first to appear in the Configuration
menu.
Configuration Menu
6. Enter the Input menu.
a. Press the Program button to enter the Input menu.
b. The Input 1 option will be the first to appear in the Input menu.
The Input Menu
7. Enter the Input 1 Menu.
a. Press the Program button to enter the Input 1 menu.
b. Sens will be the first option to appear in this menu.
Sens Option
8.
Use the arrow keys to scroll to the “OFFS” Offset parameter.
a. Scroll to the Offset calibration parameter to enter the offset.
Offset Parameter
9. Enter the temperature offset.
a. Press the Program button unlock the Offset parameter and place it in
its blinking adjustable mode.
b. Use the arrow keys to enter the offset.
-3 Offset Entered