OPERATION (CONTINUED)
OTL
A
CTIVATION
D
URING
N
ORMAL
O
PERATIONS
The following may be taking place if the OTL system activates during normal operations:
A user has set the Over Temperature Limit below the current set point for either an active
heating profile or single constant temperature set point.
An external heat source or heat source inside the chamber is causing the chamber air
temperature to spike.
The temperature controller and display unit or its sensor probe has failed, and must be
replaced in order to maintain safe operation of the oven.
If the OTL activated during normal operations adjust it very slightly clockwise to increase the
setting. If the OTL continues to interrupt heating of the oven chamber and there no obvious external
sources of nearby heating (autoclaves, another oven), depower the oven and allow the oven
chamber to cool before opening the oven door or troubleshooting.
P
ROGRAMMING A
H
EATING
P
ROFILE
Please see the
Operation and Programing JUMO dTRON Guide
included with the oven for how to
program, launch, and abort a heating profile.
Содержание SM10HP-2
Страница 17: ...17 P a g e GRAPHIC SYMBOLS CONTINUED This page left blank ...
Страница 20: ...20 P a g e OPERATION CONTINUED This page left blank ...
Страница 27: ...27 P a g e OPERATION CONTINUED This page left blank ...
Страница 33: ...33 P a g e OPERATION CONTINUED This page left blank ...
Страница 43: ...43 P a g e REPLACEMENT PART LIST This page left blank ...
Страница 44: ......