OPERATION (CONTINUED)
S
ET THE
O
VER
T
EMPERATURE
L
IMIT
Note:
Test the OTL system for functionality at least once per year.
This procedure sets the Over Temperature Limit heating cutoff to approximately 5
˚C above the
current chamber temperature.
If the OTL sporadically activates after setting the control, you may turn the dial very slightly to the
right (clockwise).
Set the OTL Cutoff Temperature
Example
1. Turn the
Set Over Temperature Limit
control dial clockwise to the
maximum position, if it is not already set to maximum.
2. Set the oven to run at either your single set point application
temperature, or a set point equal to the highest temperature of your
application heating profile.
3. Wait 30 minutes after the oven has achieved the set point for the
chamber to stabilize, before setting the OTL
4. Turn the Over Temperature Limit control dial counterclockwise
until the red Over Temperature Limit Activated light illuminates.
5. Slowly turn the dial clockwise until the Over Temperature Limit
Activated light turns off. Stop turning the control.
a. The Over Temperature Limit is now set approximately 5
˚C
above the current chamber temperature.
6.
Optional
: You may turn the dial slightly to the left to bracket in
closer to the set point temperature. This sets the OT Limit nearer
to the current chamber temperature.
7. Leave the OTL dial set just above the activation point.
Содержание SM10HP-2
Страница 17: ...17 P a g e GRAPHIC SYMBOLS CONTINUED This page left blank ...
Страница 20: ...20 P a g e OPERATION CONTINUED This page left blank ...
Страница 27: ...27 P a g e OPERATION CONTINUED This page left blank ...
Страница 33: ...33 P a g e OPERATION CONTINUED This page left blank ...
Страница 43: ...43 P a g e REPLACEMENT PART LIST This page left blank ...
Страница 44: ......