13 |
P a g e
INSTALLATION (CONTINUED)
P
OWER
S
OURCE
R
EQUIREMENTS
When selecting a location for the oven, check that
each
of the following requirements are satisfied:
The oven must be positioned so that all operators have access to the power feed
disconnect in case of emergencies.
o
The Disconnect must be in close proximity to the equipment and within easy reach
of the operator.
o
The Disconnect must be marked as the disconnecting device for the equipment.
The power source must be
single (1) phase
and
protective earth grounded
.
The power source must conform to all national and local electrical codes.
The power source voltage and ampere must match the power requirements listed on the
oven data plate (located on the back of the unit, beneath the power feed inlet).
o
These ovens are intended for a 220 - 240 volt, 50/60 Hz applications at the
following amperages:
Model
Amperage
Model
Amperage
SMO10HP-2
26
SMO34HP-2
50
SMO14HP-2
26
SMO38HP-2
50
Supplied voltage must not vary more than 10% from the data plate rating
. Damage to
the oven may result if supplied voltage varies more than 10%. Use a separate circuit to
prevent loss of the unit due to overloading or circuit failure.
A switch or circuit-breaker must be used in the building installation
to protect against
overcurrent conditions.
o
The required circuit-breakers are: SMO10HP-2 30 amps; SMO14HP-2 30 amps;
SMO34HP-2 60 amps; SMO38HP-2 60amps.
Power Exhaust Outlet Fuses
The oven is also provide with a pair (2) of
2 amp fuses
installed adjacent to the external power
outlet on the back of the unit. The outlet is solely intended to power an optional power exhaust
blower attached to the oven exhaust vent. The fuses protect against overcurrent conditions related
to the operation of the outlet and attached blower. If one fuse blows, the outlet will depower. The
cause of a blown fuse should be determined prior to replacing it.
These fuses do not provide protection against overcurrent events associated with major
components of the oven
. Overcurrent protection for the oven must be set up in the location power
supply external to the unit. See the circuit breaker requirements above.
Содержание SM10HP-2
Страница 17: ...17 P a g e GRAPHIC SYMBOLS CONTINUED This page left blank ...
Страница 20: ...20 P a g e OPERATION CONTINUED This page left blank ...
Страница 27: ...27 P a g e OPERATION CONTINUED This page left blank ...
Страница 33: ...33 P a g e OPERATION CONTINUED This page left blank ...
Страница 43: ...43 P a g e REPLACEMENT PART LIST This page left blank ...
Страница 44: ......