USER MAINTENANCE (CONTINUED)
Continued on next page
Temperature Calibration
1. Once the chamber has stabilized with no fluctuations of ±
0.2˚C or greater
compare the reference temperature device and chamber temperature
display readings.
a. If the readings are the same, or the difference between the two (2)
falls within the acceptable range of your protocol, the display is
accurately showing the chamber temperature.
The Temperature
Verification procedure is now complete
.
b. See step 2 if a difference falls outside the acceptable range of your
protocol.
Reference Device
Home Page
2. The display requires calibration.
a. The difference (also known as an error) between the reference
device and the display is an
offset
.
Examples of offset values:
Reference Sensor
Reading
Oven Temp.
Display
Offset
Value
152.0°C
150°C
2
147.0°C
150°C
-3
b. Note the offset value.
Reference Device
Home Page
Correcting an Offset
3. Jump to the Operations Menu.
a. Press the Program button.
Oprations Menu
4. Scroll to the Configuration Option.
a. Press the up arro
w button until the “Conf” option appears.
Configuration Option
Содержание SM10HP-2
Страница 17: ...17 P a g e GRAPHIC SYMBOLS CONTINUED This page left blank ...
Страница 20: ...20 P a g e OPERATION CONTINUED This page left blank ...
Страница 27: ...27 P a g e OPERATION CONTINUED This page left blank ...
Страница 33: ...33 P a g e OPERATION CONTINUED This page left blank ...
Страница 43: ...43 P a g e REPLACEMENT PART LIST This page left blank ...
Страница 44: ......