Animal scale functions F2.1 = 6
Operation method
In normal weighing state, place the animal on the weighing platform and its
weight must be ≥threshold value set in F2.2. Press
『
Total
』
, instrument will
collect data sampling. After sampling, the average value of sampling data
will be locked, showing X.XXX kg; animal state indicator flashes. Press
『
』
to print; press
『
Hold
』
or
『
Total
』
to quit the interface.
●
8
、
Prompt Message of Instrument
The instrument has extremely high stability and reliability, thus is not easy to
have error in general situation. Once an error occurs, please make clear the error
first and observe whether the instrument still has error after power-on. Do not
hurry to repair the scale body or instrument. Repair the instrument according to
error code of the instrument as possible
No.
Symbol
Analysis
Treatment Method
1
【
】
【
】
Unable to reset after
startup
1、
Determine it is no-load state
in startup;
2、
Make zero calibration again.
2
【
】
The weighed object is over
full range for 9d
Reduce weight on weighing
platform
3
【
】
The weighed object is
below 0 for 5d
Press
『
Zero
』
to reset
4
【
】
【
】
Out of zero clearing range
Check whether the weighing
platform has weight.
Remove weight.
5
【
】
Invalid operation
6
【
】
EEPROM checksum and
error
Press
『
』
reprint factory
value. Start up again. If the
information
occurs
again,
return to factory for repair.
Please calibrate the scale again
if the situation does not occur;
Attention:
this place is provided
with
all
parameters
of