8
0.00 ---- 2 decimal places
0.000 ---- 3 decimal places (default value)
0.0000 ---- 4 decimal places
F1.3 Number of Divisions
Selectable parameters: 1 (default value), 2, 5, 10, 20, 50
F1.4 Calibration Unit
Selectable parameters: 0 ---- kg (default value)
1 ---- lb
F1.5 Gravitational Acceleration
Selectable parameters: 9.70000~9.99999. Default value = 9.79455.
F1.6 Null-point Calibration
【
】
Keeping empty the scale
Remove the weights on the weighing platform
to guarantee the scale is in the empty state.
Press the
『
ON/OFF
』。
key and the indicator will
display
【
】,
The displayed digits will
reduce slowly until the indicator displays
【
】,
In the end it will display
【
】
for one second, which indicates the end of
null-point calibration.
F1.7 Load-point Calibration
【
】
Loading weights
Load weights on the weighing platform to ensure
that 10% of full-scale value ≤ weight of weights
≤ full-scale value, and then press the
『
ON/OFF
』
key to start the next step.
【
】
Entering the same weight value as that of the
loaded weights.
Entering the same weight value as that of the
loaded weights, please press the
『
ON/OFF
』
key
after the scale becomes stable, and the meter
will display [
]. After that, the displayed
digits will reduce slowly until the meter displays