9
[
]. In the end it will display [
] for
one second, which indicates the end of
null-point calibration.
F1.8 Automatic Null Tracking
Selectable parameters: OFF, 1 d, 2 d, 3 d, 5 d (default value)
F1.9 Automatic Reset Range at Startup
Selectable parameters: OFF, 2 %, 10 %, 20 % (default value)
F1.10 Button Reset Range
Selectable parameters: OFF, 2 %, 10 % (default value), 20 %
F1.11 Digital Filter
Selectable parameters:
0 ---- Mild Filtering
1 ---- Moderate Filtering (default value)
2 ---- Severe Filtering
F1.12 Steady Range
Selectable parameters:
1 d, 2 d, 3 d (default value)
F1.13 Overload Display Range
Selectable Parameters: 9d, 5% (default value), 10%, 20% ±120%
F1.14 F1 Menu Protection
Selectable Parameters: 0 ---- Enter F1 menu by keyboard
operation
1 ---- Enter F1 menu by pressing the calibration button
F1.15 Strong filter mode
Selectable Parameters:0 ---- Not enabled
1 ---- start using
F1.16 Restoring Factory Default
Set the parameters within F1~F4 as the defaults, which can’t
impact the parameters of standard scale.
F2
Application Function Setting
F2.1 Function Selection
Selectable Parameters: 0 ---- ×10 functions