7
Function key->
『
Func
』
Long key → Enter into setting menu.
Flicker bit is on the
right.
Tare key->
『
Tare
』
Short key →gross weight turns to net weight.
Indication lamp of net weight “Net” is on.
Conduct skin removal operation for multiple
times.
Digit
flicker
position reduces.
ZERO Key->
『
Zero
』
Gross weight state resets weight. When the
scale is in net weight, dynamic state, saving
state and out of resetting range, zero clearing
operation is invalid.
In setting, digit of
flicker position
increases. In
adjustment of display,
accumulation is
cleared.
Short key → Start up
。
Long key → shut down.
Confirm operation,
to save setting
data.
6
、
Parameter Setting
Setting entry:
Press the
『
Func
』
button on the operating panel in the state of
normal weighing.
If F1.14 = 0, you can set all the parameters within F1~F5.
If F1.14 = 1, you can only set all parameters within F2~F5.
If F1.14 = 1 and you need to set the parameters within F1 menu, you
can press the calibration switch button until the F1 menu is entered.
F1
Parameter Setting
F1.1 Measuring Range
Selectable parameters: 3~200000 (default value: 6)
F1.2 Decimal Places
Selectable parameters: 0 ---- no decimal point
0.0 ---- 1 decimal place