12
F5.1 Key test
Instrument display
【
】
, press
『
』、
『
Zero
』、
『
Tare
』、
『
Gross
』、
『
lb/kg
』、
『
Total
』
in order, and the instrument displays
【
】、【
】、【
】、【
】、【
】、
【
】,
press
『
Hold
』
to quit key test.
F5.2 Display screen test
All strokes of meter display will have self-inspection, to observe
whether there is lacks of strokes.
Press
『
Hold
』
or
『
ON/OFF
』
to quit test of display screen.
F5.3 Display current internal code
The display will show internal code of current instrument after
smoothing. Press
『
Hold
』
or
『
ON/OFF
』
to quit the interface.
。
7
、
Function Description
● ×10 function F2.1 = 0
Operation
In normal weighing mode, press
『
Total
』
on the operation panel once,
the display accuracy will promote for 10 times and icon ×10 gets on at
the same time; press the key again it returns to normal weighing mode
and the icon of ×10 gets off.
Note: 1. In net weight status, it’s forbidden to change over to ×10 function status;
2. If there have been 4 decimal places, it’s forbidden to change over to ×10 function
status;
3. Under ×10 function status, both serial output and printing output are forbidden.
4. Under ×10 function status, it’s forbidden to tare.
● Weight hold function F2.1 = 1
How to operate In normal weighing mode, press
『
Hold
』
on the operation
panel once, the weight presently displayed can be locked and “Hold”
indicator gets. Weight Hold operation is valid only when the displayed
weight is ≥F2.2 set value; otherwise, it displays invalid operation with
【
】
for one second and the scale goes back to normal weighing
mode.