7-2.
Calibration Process
Error
Cause
Treatment
Err 01
When Max.capacity/digit value is over
20,000
Re-input the Max. Capacity, less than 20.00
(Max. Capacity / Digit)
Err 04
Standard weight value is over than
Max. Capacity
Re-input Standard weight value with Number
keys, under Max. Capacity
Err 05
Standard weight value is less than
10% of Max. Capacity
Re-input Standard weight value with Number
keys, more than 10% of Max. Capacity
Err 06
1. Amp. Gain is too big
2. Sig+ and Sig- wire connection error
3. Test weight is not loaded
Check standard weight’s weight with set value.
If there is difference between set value and real
weight, please re-input the value
(set value is too small)
Err 07
1. Amp. Gain is too small
2. Sig+ and Sig- wire connection error
3. Test weight is not loaded
Check standard weight’s weight with set value.
If there is difference between set value and real
weight, please re-input the value
(set value is too big)
Err 08
Under “F-function” model, set value is
“N.A”
Check the correct value and re-input
Err A
When there is continuous vibration on
the weighing part,, indicator cannot
process calibration any more.
- Find vibration cause and remove
- Load cell check
- Load cell cable and connecting condition
check