60
7-3.
Digital Weighing Indicator
Error
No.
Display
Cause
Treatment
No.1
“CELL-Er”
or
“OVER”
1. Load cell Error
2. Load cell cable Error
3.Load cell connection Error
4. A/D Board Error
5.If Analogue value
is over 1,040,000.
※ When weigh “-“ value,
If it is over set max capa, “OVER” is
displayed.
Ex) Even though set max capa is “100” and
it is over “-100”, “OVER” is displayed.
1. Under “TEST” mode 1, check
analogue value. If you cannot get
any analogue value or there is no
change although adding load,
please check load cell, load cell
cable, connection conditions first.
2. Replace another load cell, and check
the indicator condition. If you have
same problem, please replace new
indicator and check A/D board
error.
3. Try to connect the indicator’s A/D
with the other indicator.
4. Check the power and connection of
terminal.
No.2
“UnPass”
1. Power is ON, when some materials
are on weighing part.
※ Under “Normal Mode”, if there are
more than 20% loading of Max.
capacity, “Un-Pass” display will be
appeared and indicator will stay
until removing the load.
※Setting Back-up mode it can
memory empty value, and it
becomes set value without
displaying” Un-pass”)
1. If you set “Normal Mode”, please
check weighing part empty or not
before turn on the power. If there
are some materials in/on weighing
part, please remove those materials
and turn on the power.
2. Please try to set F02-01(Back-up)
mode so that the indicator can
remember first empty value.
No,3
“SET”
When Power is on, “SET” displays.
It means
EEPROM has some problem.
Please contact the distributor or Head
Office.
No,4
“halt”
H/W has some problem.
No.5
“T-Err”
The dead Battery
※ Under “CELL-Er”, Zero key, Tare key, Hold key and print key will not be activated.