- 31 -
When
“
Weighing Judgment Finish relay
”
output, Judgment data will be saved and printed automatically.
Under this setting,
“
Saved data
”
and
“
Printed data
”
is same. So, this setting is recommended for
Automatic weighing process.
Weight-Back up selection
●
0
Normal
Mode
F02
1
Weight Back up Mode
Normal Mode
※
When the power is off , “Zero value” will be out and will display wrong weight value after power is
recovered.
Also, if there is material over 10% of max. capacity weight, the
“Un-Pass”
display will be appear and
you can not process weighing job.
If there is material less than 10% of max. capacity weight on the scale, indicator will initialize
with
”Pass”
display, and display
“0”
whatever the real weight is. Under this case, remove the material
and press
“Zero”
key and input new
“Zero”
value. Then, new Zero value will be memorized and you
can process weighing job.
※
Weight Back up Mode
Under this mode, SI 4410 memory
“Zero” value
of weighing part. So, when the power is recovered
SI4410 can display material weight.
If you remove and turn on the power, please press “Zero” key and re-memorize Zero value.
Motion Band Range setting
F03 2
1
∫
9
This is set “Steady” acceptable range of weighing part.
If there is vibration on weighing part, you can set this function and
reduce the vibration effect on weighing process.
1 : Weak vibration
∫
9 : Strong Vibration
This f
※
unction is compensate vibration effect on weighing part with acceptable range setting.
During the fixed time period(F06 setting), there is smaller weight variation than Motion band range, due
to vibration, Indicator will display “Steady”, if there is larger weight variation than Motion band range,
due to vibration, Indicator will its weight and “Steady” condition is broken and find new “Steady” point.
If there are much vibration effect, please set with large set value.