- 25 -
Under this step, input Max. Output rate(mV) of load cell.
The Output rate is stated on “Test report” or “Label” and please input this value with No. keys.
Normally, the Output rate will be 4digits under Zero, then input 4digits and input “0” additionally and
make full 6digits.
After input the value, press
key to make calculation.
After finishing calculation,
Calculated “Span value” will be
display.
Caution
※
Due to some variation between
“State output rate”
and
“Real Output rate”
of load cell, there might
be some weight difference after finishing calibration.
If you want to make more precise weighing process, please measure real output rate of load cell and
input the measured value.
Then the weight measurement will be more precise than before.
Step No.6 – End calibration process and Calibration S/W OFF
Calibration process completed.
Push Calibration Lock S/W and tight Cal. Bol.
Now, the Simulation Calibration is done.
Press
key to save all calibration process and Off the
“CAL LCK S/W”
and fasten the
Calibration Bolt.