LISST-Tau
User’s Manual
Operation Details
Page 18
Date: 2010-01-01T40:26:00
Serial number: 1013R
Firmware version: 1.46
Calibration date: 2021-07-29T13:10:30
Temperature Corr. Coeff.: 1.00 0.00
0.00 0.00 0.00 0.00
Transmission Cal.: 1.00
Temperature Cal.: 2.01
Temperature Func. Cal.: 1.00
S
Set Clock
Syntax:
S [yy]yy/mm/dd HH:MM:SS
Description:
Set internal Real Time Clock. Where
yy
= year (may include the century,
however, century is ignored),
mm
= month,
dd
= day of month,
HH
= hour
(00 to 23),
MM
= minute,
SS
= second.
For example, September 29, 2019 at 10:12:14 would be 19/9/29 10:12:14.
Sending S only will display the current clock value.
Example:
input:
S
output:
2000-01-01T00:03:31
input:
S 19/12/31 11:59:59
output:
RTC set to 2019/12/31 11:59:59
Z
Start Clean Water Re-Calibration (Baseline Measurement)
Syntax:
Z
Description:
Re-calibrate the Beam-C and Transmission measurements to be relative
to current values. Instrument is powered up for 30 seconds before an
average is obtained to compute the calibration. Ctrl-C to cancel.
Example:
input:
Z
output:
Warming for 30 seconds.
30 29
…
3 2 1
Calibrating with 10 samples. Press <Ctrl>C to cancel.
LTAU1013R 2021-09-02T15:28:23 2136 31647 48137 16490
…
LTAU1013R 2021-09-02T15:28:24 2137 31640 48122 16482
…
LTAU1013R 2021-09-02T15:28:26 2139 31650 48132 16481
…
LTAU1013R 2021-09-02T15:28:27 2140 31641 48136 16495
…
LTAU1013R 2021-09-02T15:28:29 2142 31648 48126 16477
…
LTAU1013R 2021-09-02T15:28:31 2143 31642 48135 16492
…
LTAU1013R 2021-09-02T15:28:32 2145 31645 48127 16481
…