LISST-Tau
User’s Manual
Operation Details
Page 16
E.
Serial Commands
Command Summary
NOTE: Command are shown in upper case for clarity but are not case-sensitive.
DEFAULTS
Restores to standard factory operating mode and settings
(without affecting calibration)
DS
Display instrument configuration data such as Serial Number
and calibration information.
GO
Start continuous data output
H
Display Help on available commands
S
Set Clock using format yy/mm/dd HH:MM:SS
Z
Start clean water re-calibration
Command Details
DEFAULTS
Restore standard factory operating setup
Syntax:
DEFAULTS
Description:
The LISST-Tau has various settings and modes, not described in this
manual, for troubleshooting or special applications. The DEFAULTS
command restores the LISST-Tau to its factory-standard operating
parameters for operation as described in this manual. It does NOT affect
the user calibration.
Example:
input:
DEFAULTS
TAU:>defaults
Reset to factory default settings (will not affect
calibration)? [y|n] y
Set to factory defaults.