LISST-
Tau User’s Manual
Operation Details
Page 17
GO
Start Continuous Data Output
Syntax:
GO
Description:
Starts continuous data output at 1 sample per second. Output will be in
format described in Appendix B: Data Output Format. Data output can be
stopped by pressing CTRL-C.
Example:
input:
D
output:
LTAU2 2000-01-01T00:00:49 38932 39434
1.0015 0.9355 0.4442 26.9 26.3 24.7
11.29 1.31 2019-12-23T11:39:04 1.08139
21.79442 1.01021
H
Display Help
Syntax:
H
Description:
Display list of commands.
Example:
input:
H
Output:
LISST-Tau 1.46 (Aug 24 2021 09:57:46)
-----------------------------------------------------
DS Display status.
GO Start continuous data
collection.
S [yy/mm/dd HH:MM:SS] Read or set instrument clock.
Z [dddd] Determine the clean water
calibration. Number of samples [dddd] is optional
(default 10 samples).
DS
Display Status / Configuration
Syntax:
DS
Description:
Display current instrument configuration and calibration parameters
Example:
input:
DS
output: