5.4 Wake up / power up LISST-HOLO
If the system is still powered up but asleep, move the switch to the 1 position
and then quickly back to the 0 position. This will wake up the instrument.
If the system is disconnected from power, make sure the switch is at the 0
position, then power it back on. The LISST-HOLO will boot up and be ready in
approximately 30 seconds.
When the green LED is flashing every few seconds the instrument is awake and
ready to start sampling.
5.5 Start Data Collection
To start the data collection program, move the Switch to the 1 position.
The LED on the endcap will change from green to red indicating that sampling
has begun.
It will blink slowly until the start condition is reached and it will be on
continuously during the acquisition of a hologram.
It will return to green when the stop condition has been reached.
5.6 Stop Data Collection
Data collection will continue until the selected Stop Condition is reached. To
Stop the data collection at any time, move the switch to the 0 position. Within a
few seconds the LED on the endcap should change from Red to Green.
If the data collection has already stopped the LED will be blinking green
regardless of the switch position. However, the switch should be always be
placed in the 0 position when sampling is complete.
5.7 Downloading Raw Holograms
The raw holograms are saved to the External Memory Module if it is installed.
Otherwise the holograms are stored on the internal drive of the LISST-HOLO.
The offload procedure is different for those two scenarios and will be described
in the following.
Downloading from EMM
Raw holograms stored on the EMM can be downloaded using the provided
USB cable.
Remove the EMM from the instrument. Leave the locking ring attached to the
connector on the instrument. Plug the mating underwater connector of the USB
cable onto the EMM and then plug the USB end into your PC. The USB flash
drive inside the EMM should now appear on your PC. You can now transfer
files as you would from any USB flash drive and subsequently delete the
holograms you no longer wish to keep on the EMM.
See section 5.8 for instructions on how to view the raw (unprocessed)
holograms using OpenSeeIt.
See section 4 for instructions on how to process the holograms.
Downloading and deleting from LISST-HOLO
internal drive (EMM not installed)
To download holograms from the internal drive you will need an FTP utility like
Filezilla (
). When you have started the FTP application,
you will need to specify a host, username and password.
The host is the IP address of the LISST-HOLO, typically
192.168.0.150
unless
you have changed it to another IP address (see section 8.3).
Leave the username blank
The password is
anonymous
Содержание LISST-HOLO
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 10: ...10 LISST HOLO User s Guide...
Страница 14: ...14 LISST HOLO User s Guide...
Страница 20: ...20 LISST HOLO User s Guide...
Страница 57: ...LISST HOLO User s Guide 57...
Страница 64: ...64 LISST HOLO User s Guide...
Страница 70: ...70 LISST HOLO User s Guide...