4 Troubleshooting
SenTec Digital Monitoring System
69
4.2.5
Docking Station specific troubleshooting
Table 10
Docking Station specific troubleshooting
Problem
Possible Cause(s)
Recommended action(s)
P0500
Docking
Station door
does not
close
properly
Sensor cable
jammed in the door
Adjust cable and close the Docking Station
door properly.
Sensor not inserted
properly in the
Docking Station
door.
Insert the sensor properly and close the door.
Mechanical defect
of the Docking
Station door
Contact qualified service personnel.
Mechanical defect
of the Docking
Station door
Replace Docking Station door according to
section
Gasket not placed
correctly
Place gasket correctly in the chamber in the
Docking Station according to
section
“Replacement, Docking Station Gasket”.
Docking Station
door mechanically
defective
Replace Support for Docking Station Door
according to the
Repair Manual for the SDMS.
To be performed by SenTec authorized
service technicians only
.
P0501
Message
‘Docking
Station Fault’
is displayed
Contaminations
(e.g. gel residues)
disturb the gas flow
Clean Docking Station and SenTec TC Sensor
and then insert the sensor into the Docking
Station.
If problem persists contact qualified service
personnel.
Note: Refer to the
Instruction Manual for the
SDMS.
Docking Station
defective
Contact SenTec authorized service personnel.
Docking Station
Module defective
Replace Docking Station Module according to
the
Repair Manual for the SDMS.
To be performed by SenTec authorized
service technicians only
.
P0502
Message ‘Gas
leak in DS’ is
displayed or
a gas leak is
Sensor not properly
positioned in
Docking Station
during last
calibration.
Verify that the sensor is clean and is properly
placed into the holder at the inside of the
Docking Station door and that the door
properly closes (refer to the
Instruction
Manual for the SDMS).
Содержание SDM
Страница 1: ...For the SenTec Digital Monitoring System Software Version SMB V08 02 MPB V06 02 ...
Страница 6: ......
Страница 28: ...3 Maintenance Safety Functionality Tests HB 005615 j 22 Figure 6 Examples PCO2 and PO2 Slope Test ...
Страница 84: ...5 Repairs of the SDM HB 005615 j 78 Figure 17 Positioning of the Docking Station Gasket ...
Страница 93: ...8 Appendix SenTec Digital Monitoring System 87 HB 005615 j Date of release 02 2019 ...