3 Maintenance/Safety & Functionality Tests
HB-005615-j
42
3.4
Digital Sensor Adapter Cable
3.4.1
Cleaning and Disinfection
Clean or disinfect the `Digital Adapter Cable´ as often as required per institutional
ordinances (recommended weekly). For cleaning and disinfection instructions refer to
the
Technical Manual for the SDM (HB-005752).
3.4.2
Visual Inspection
Check the Digital Sensor Adapter Cable for damages of the isolation. Check the Digital
Sensor Adapter Cable connectors for damages.
If you detect damages, the Digital Sensor Adapter Cable needs to be replaced by a
new Digital Sensor Adapter Cable.