4 Troubleshooting
HB-005615-j
56
Problem
Possible Cause(s)
Recommended action(s)
Clean the sensor as described in
section
“Cleaning / Inspecting without Membrane”.
Then calibrate the sensor.
SenTec TC Sensor
defective
If the problem persists the SenTec TC Sensor
is defect. Replace it with a new one.
Calibration Module
defective
Replace the Calibration Module according to
the
Repair Manual for the SDMS.
To be performed by SenTec authorized
service technicians only.
SDM defective
Refer to
section
troubleshooting ” to further evaluate the
problem.
P0304
PCO2 slow
(too slow) /
Message
‘Sensor
problem 11,
14’
OR
PO2 slow
(too slow) /
Message
‘Sensor
problem 74’
Membrane related
(defective or dried
up electrolyte)
Change the membrane :
-
if it is damaged or missing, has a loose fit,
or if there is trapped air or dry electrolyte
under the membrane
-
sensor calibration takes almost 14
minutes or cannot be completed within 14
minutes
Note: If the PCO2 calibration takes 14
minutes the message ‘PCO2 slow’ displays
and PCO2 subsequently will be flagged as
questionable.
Note: If the PCO2 calibration cannot be
finished within 14 minutes the message
‘Sensor problem 11’ displays and PCO2
subsequently will be flagged as invalid.
Note: If the calibration cannot be finished
within approx. 30 minutes – due to failed
extended calibration – the message ‘Sensor
problem 14’ displays and PCO2 subsequently
will be flagged as invalid.
Note: If the PO2 calibration cannot be
finished within 14 minutes the message
‘Sensor problem 74’ displays and PO2
subsequently will be flagged as invalid.
PCO2 electrode
and/ or white PO2
sensing spot
contaminated
If after having changed the sensor membrane
the calibration still takes almost 14 minutes or
cannot be completed within 14 minutes
contact qualified service personnel to clean
the sensor without membrane.
Clean the sensor as described in
section
“Cleaning / Inspecting without Membrane”.
Then calibrate the sensor.
Содержание SDM
Страница 1: ...For the SenTec Digital Monitoring System Software Version SMB V08 02 MPB V06 02 ...
Страница 6: ......
Страница 28: ...3 Maintenance Safety Functionality Tests HB 005615 j 22 Figure 6 Examples PCO2 and PO2 Slope Test ...
Страница 84: ...5 Repairs of the SDM HB 005615 j 78 Figure 17 Positioning of the Docking Station Gasket ...
Страница 93: ...8 Appendix SenTec Digital Monitoring System 87 HB 005615 j Date of release 02 2019 ...