3 Maintenance/Safety & Functionality Tests
SenTec Digital Monitoring System
23
Slope Test Control Table CO2 – in mmHg
(after 5% and 30% CO2 calibration)
Pcal = value after calibration with 8%CO2.
Table is only valid for sensor temp. of 42°C.
P
atm
Pcal 5_min
5_ideal
5_max
30_min
30_ideal
30_max
600
33.5
19.0
19.8
21.1
123
134
144
605
33.8
19.2
20.0
21.3
124
135
145
610
34.1
19.4
20.2
21.5
125
136
146
615
34.4
19.6
20.4
21.7
127
137
148
620
34.7
19.8
20.6
21.9
128
138
149
625
35.0
19.9
20.8
22.1
129
139
150
630
35.3
20.1
21.0
22.3
130
140
151
635
35.6
20.3
21.1
22.5
131
142
152
640
35.9
20.5
21.3
22.7
132
143
154
645
36.2
20.7
21.5
22.9
133
144
155
650
36.5
20.9
21.7
23.1
134
145
156
655
36.8
21.0
21.9
23.3
135
146
157
660
37.1
21.2
22.1
23.5
136
147
159
665
37.4
21.4
22.3
23.7
137
148
160
670
37.7
21.6
22.5
23.9
138
150
161
675
38.0
21.8
22.7
24.1
139
151
162
680
38.3
21.9
22.9
24.3
140
152
163
685
38.6
22.1
23.0
24.5
141
153
165
690
38.9
22.3
23.2
24.7
142
154
166
695
39.2
22.5
23.4
24.9
143
155
167
700
39.5
22.7
23.6
25.1
144
156
168
705
39.8
22.9
23.8
25.3
145
157
170
710
40.1
23.0
24.0
25.5
147
159
171
715
40.5
23.2
24.2
25.7
148
160
172
720
40.8
23.4
24.4
25.9
149
161
173
725
41.1
23.6
24.6
26.1
150
162
174
730
41.4
23.8
24.7
26.3
151
163
176
735
41.7
24.0
24.9
26.5
152
164
177
740
42.0
24.1
25.1
26.7
153
165
178
745
42.3
24.3
25.3
26.9
154
167
179
750
42.6
24.5
25.5
27.1
155
168
181
755
42.9
24.7
25.7
27.3
156
169
182
760
43.2
24.9
25.9
27.5
157
170
183
765
43.5
25.1
26.1
27.7
158
171
184
770
43.8
25.2
26.3
27.9
159
172
185
Содержание SDM
Страница 1: ...For the SenTec Digital Monitoring System Software Version SMB V08 02 MPB V06 02 ...
Страница 6: ......
Страница 28: ...3 Maintenance Safety Functionality Tests HB 005615 j 22 Figure 6 Examples PCO2 and PO2 Slope Test ...
Страница 84: ...5 Repairs of the SDM HB 005615 j 78 Figure 17 Positioning of the Docking Station Gasket ...
Страница 93: ...8 Appendix SenTec Digital Monitoring System 87 HB 005615 j Date of release 02 2019 ...