3 Maintenance/Safety & Functionality Tests
HB-005615-j
20
3.1.9
PCO2 / PO2 Slope Test
Tools:
Sensor Tester Unit (alternatively an SDM can be used)
V-STATS Software (optional, to enable calibration curve on SDM)
interface online log software, e.g. “Hterm” freeware available in
the internet (optional, for easier on-line logging of calibration
values)
Note: For a V-Sign™ Sensor only the PCO2 Slope Test must be performed. For an
OxiVenT™ Sensor both, PCO2 and PO2 Slope Test, must be performed.
Note: The Slope Test can also be performed using a second SDM instead of a
Sensor Tester Unit. Also, a single SDM can be used (requiring changing gas
bottles after each step). If you use one single SDM, make sure that
the correct gas bottle with the correct gas concentration (with 8%
CO2, 12% O2) remains in the SDM at the end of the test.
Note: Prior performing the Slope Test please ensure that the sensor is properly
stabilized, i.e. stored for at least 4 hours in the Docking Station with the
monitor switched on. For an overview about the Slope Test please refer to
Examples: PCO2 and PO2 Slope Test.
Note: After the test ensure to reset all parameters and settings to the ones
originally selected by your customer
1.
You may optionally activate the display of the calibration curve by using “V-
STATS” (PC-Software for the SDM).
2.
Verify proper function of the SenTec Sensor Tester (ST-DSU). In particular verify
the integrity of the Docking Station. Verify the expiration date of the test gas
bottles.
3.
Power-cycle the SDM before starting the Slope Test (i.e. turn the SDM off and on
again to ensure that the next calibration is a so-called `Initial Calibration´, which
calibrates both, the PCO2 and PO2 channel).
4.
Verify that the sensor SET temperature is 42°C or else set the sensor temperature
to 42°C (for V-Sign™ Sensor and OxiVenT™ Sensor), the ‘Severinghaus
Correction Mode’ is set to AUTO and that the local atmospheric pressure is at
least 600 mmHg (80 kPa) (see respective icon). Note down the atmospheric
pressure value. If you are using an OxiVenT™ Sensor ensure that the PO2-
parameter is enabled (e.g. select Neo-mode and enable PCO2/PO2 parameters
and set sensor SET-temperature to 42°C).
5.
Insert the sensor into the Docking Station of the SDM. Sensor calibration starts
automatically. If not: Use the menu-function ‘Measurement Settings / PCO2
Settings / Calibrate sensor’ to manually activate the sensor calibration (note: In
case of an OxiVenT™ Sensor this function will initiate a PCO2 and PO2
calibration). After the calibration (with 8% CO2, 12% O2) is finished please wait
6 minutes (this is to ensure that no gas leak is present in the Docking Station of
the SDM).
Содержание SDM
Страница 1: ...For the SenTec Digital Monitoring System Software Version SMB V08 02 MPB V06 02 ...
Страница 6: ......
Страница 28: ...3 Maintenance Safety Functionality Tests HB 005615 j 22 Figure 6 Examples PCO2 and PO2 Slope Test ...
Страница 84: ...5 Repairs of the SDM HB 005615 j 78 Figure 17 Positioning of the Docking Station Gasket ...
Страница 93: ...8 Appendix SenTec Digital Monitoring System 87 HB 005615 j Date of release 02 2019 ...