3 Maintenance/Safety & Functionality Tests
HB-005615-j
26
Slope Test Control Table PO2 – in kPa
(after 6% and 21% O2 calibration)
Pcal = value after calibration with 12% O2
Table is only valid for sensor temp. of 42°C.
P
atm
Pcal 6_min
6_ideal
6_max
21_min
21_ideal
21_max
80.0
9.6
4.6
4.8
5.0
15.8
16.8
17.8
80.7
9.7
4.6
4.8
5.1
15.9
16.9
18.0
81.3
9.8
4.6
4.9
5.1
16.0
17.1
18.1
82.0
9.8
4.7
4.9
5.2
16.2
17.2
18.3
82.7
9.9
4.7
5.0
5.2
16.3
17.4
18.4
83.3
10.0
4.7
5.0
5.2
16.4
17.5
18.5
84.0
10.1
4.8
5.0
5.3
16.6
17.6
18.7
84.7
10.2
4.8
5.1
5.3
16.7
17.8
18.9
85.3
10.2
4.9
5.1
5.4
16.8
17.9
19.0
86.0
10.3
4.9
5.2
5.4
17.0
18.1
19.1
86.7
10.4
4.9
5.2
5.5
17.1
18.2
19.3
87.3
10.5
5.0
5.2
5.5
17.2
18.3
19.4
88.0
10.6
5.0
5.3
5.5
17.4
18.5
19.6
88.7
10.6
5.1
5.3
5.6
17.5
18.6
19.7
89.3
10.7
5.1
5.4
5.6
17.6
18.8
19.9
90.0
10.8
5.1
5.4
5.7
17.8
18.9
20.0
90.7
10.9
5.2
5.4
5.7
17.9
19.0
20.2
91.3
11.0
5.2
5.5
5.8
18.0
19.2
20.3
92.0
11.0
5.2
5.5
5.8
18.2
19.3
20.5
92.7
11.1
5.3
5.6
5.8
18.3
19.5
20.6
93.3
11.2
5.3
5.6
5.9
18.4
19.6
20.8
94.0
11.3
5.4
5.6
5.9
18.6
19.7
20.9
94.7
11.4
5.4
5.7
6.0
18.7
19.9
21.1
95.3
11.4
5.4
5.7
6.0
18.8
20.0
21.2
96.0
11.5
5.5
5.8
6.0
19.0
20.2
21.4
96.7
11.6
5.5
5.8
6.1
19.1
20.3
21.5
97.3
11.7
5.5
5.8
6.1
19.2
20.4
21.7
98.0
11.8
5.6
5.9
6.2
19.3
20.6
21.8
98.7
11.8
5.6
5.9
6.2
19.5
20.7
22.0
99.3
11.9
5.7
6.0
6.3
19.6
20.9
22.1
100.0
12.0
5.7
6.0
6.3
19.7
21.0
22.3
100.7
12.1
5.7
6.0
6.3
19.9
21.1
22.4
101.3
12.2
5.8
6.1
6.4
20.0
21.3
22.5
102.0
12.2
5.8
6.1
6.4
20.1
21.4
22.7
102.7
12.3
5.9
6.2
6.5
20.3
21.6
22.9
Содержание SDM
Страница 1: ...For the SenTec Digital Monitoring System Software Version SMB V08 02 MPB V06 02 ...
Страница 6: ......
Страница 28: ...3 Maintenance Safety Functionality Tests HB 005615 j 22 Figure 6 Examples PCO2 and PO2 Slope Test ...
Страница 84: ...5 Repairs of the SDM HB 005615 j 78 Figure 17 Positioning of the Docking Station Gasket ...
Страница 93: ...8 Appendix SenTec Digital Monitoring System 87 HB 005615 j Date of release 02 2019 ...