Troubleshooting
9
Whenever an error occurs, the following table might help to identify the cause of the error and may
provide a possible solution:
Symptom
Cause of Error
Troubleshooting
Device cannot be turned
on
Batteries incorrectly inserted Insert batteries correctly (see
)
Battery compartment is not
closed correctly
Lock the lids of the battery compartments
tightly
Empty batteries
Replace batteries
Device can be turned on,
but doesn't display an
values
Choice of incorrect
measurement mode or
measurement range
Select a more sensitive measurement
mode
Defective device
Send in the device to SENSYS for repair
Data doesn't correspond
to expectations of the
operator
Choice of incorrect
measurement mode or
measurement range
Select a measurement mode/range that
corresponds to the survey type/object to
be located (see
)
No connection can be
established between
SBL10 and PDA
Incorrect connection of
Bluetooth dongle
Check if Bluetooth dongle is plugged in
correctly
Bluetooth turned off
Turn on Bluetooth as described in
No connection of SBL10
with PDA
Connect devices as describes in
Serial interface not assigned Assign serial interface to Bluetooth
connection as described in
Incorrect configuration in
MonSX
Reconfiguration as described in
Bluetooth connection
established, but
connection is unstable
Connection disturbed due to
environmental influences
Use a cable instead of a Bluetooth
connection.
19
14
29
29
29
32