Noggin
13-Survey & Map Mode
93
13.2.12 Filter
The filter has the effect of removing flat-lying events in the data and enhancing dipping and
hyperbolic reflections usually caused by pipes and other point targets. It can also assist in seeing
very shallow targets that may be masked by the strong signals at the top of the cross-section.
The image below shows the same data with the Filter OFF and ON.
The filter is only applied to the data for display; saved data files contain raw data and are not
filtered.
The filter can be turned on and off when replaying data (
13.2.1: p.78
) so it is not necessary to
collect the survey line again in order to apply the filter.
13.2.13 Error Messages
If an error occurs during data acquisition, an error message will appear in the bottom left section
of the data acquisition screen. Note the message number, exit the program and turn off the
Digital Video Logger.
Make sure the cables are not damaged and that all cable connections are tightly secured.
Sometimes vibrations cause the cable connections to loosen just a bit and break contact and this
can cause errors. Disconnecting cables and reconnecting them may provide a better contact
and solve the problem. Also check and make sure the battery is adequately charged. Turn the
Digital Video Logger back on and try running the system again.
For more information on Troubleshooting the system, see
14: p.122
.
Содержание Noggin 100
Страница 1: ...User s Guide Copyright 2012 Sensors Software Inc 2011 00022 02 Noggin ...
Страница 2: ......
Страница 4: ......
Страница 8: ......
Страница 17: ...Noggin 2 Noggin Components 3 Figure 2 3 Noggin 100 components ...
Страница 156: ...Appendix D Instrument Interference Noggin D 2 ...
Страница 158: ...Appendix E Safety Around Explosive Devices Noggin E 2 ...
Страница 162: ...Noggin G 4 ...