![SENSIBLE MEDICAL ReDS System V2.6 Скачать руководство пользователя страница 96](http://html1.mh-extra.com/html/sensible-medical/reds-system-v2-6/reds-system-v2-6_user-manual_1243483096.webp)
Technical Information
94
Immunity Test
IEC 60601
Level
Compliance
Level
Electromagnetic Environment
Guidance
Conducted RF,
IEC 61000-4-6
3 Vrms
150 kHz to 80
MHz
6 Vrms
in ISM bands
between 150
kHz to 80
MHz
80 % AM at 1
kHz
[
V
1
] = 6 Vrms
150 kHz to 80
MHz
Recommended separation distance:
d = 0.6 P
Radiated RF,
IEC 61000-4-3
10 V/m
80 MHz to
2.7 GHz
[
E
1
] = 10 V/m
Recommended separation distance:
d = 1.2 P, 80 – 800 MHz range
d = 2.3 P, 800 – 2,500 MHz range
Proximity fields
from RF
wireless
technologies,
IEC 61000-4-3
9- 28 V/m
According to
Table 9 in IEC
60601-1-2 4
th
edition
[
E
1
] = 9- 28
V/m
According to
Table 9 in IEC
60601-1-2 4
th
edition
Recommended separation distance:
d= 30 cm