Cautions
10
CAUTIONS Continued
14. In the following cases remove the sensor vest, turn off the system, disconnect the
power supply and contact support:
Recurring device malfunction message
Device is not responding
A measurement that does not stop automatically within four (4) minutes of
starting.
15. When connecting the USB modem, to reduce risk of damage to the system, do
not touch the system's USB port as identified by the following label:
16. The use of USB accessories other than those specified may result in increased
electromagnetic interference or decreased electromagnetic immunity of the
ReDS System.
17. The ReDS System should not be used adjacent to, or stacked with, other
electronic equipment. See Table 7 and Table 8 for minimal separation distances.
18. To ensure accurate measurements make sure to read and enter the chest size
ruler value accurately.
19. The system is intended for indoor use and storage, avoid exposing to high or low
temperatures, humidity and pressure conditions (See Section C Technical
Specifications) as this may cause system malfunctions. To avoid patient mix up,
when it is not the first time the patient is being measured, make sure you are
selecting the correct patient from the list, according to the unique patient ID
and/or other patient details.
20. Stop using the device if there is a reason to suspect that it has been
contaminated.
21. Make sure to use the vest with proper disposable infection control garments to
protect patients.
22. Do not connect devices to this system’s USB port, except as instructed in this
manual.
23. Do not connect any unauthorized USB devices to the USB port. Please consult
with your institution's IT/bioengineering department to get guidance on which
USB devices are authorized, and how to use them.