Taking a Measurement
59
Wait until the measurement process ends automatically and the red chest sensor deflates.
After a successful measurement the following screen is displayed.
Figure 43: Measurement Results
Remove the sensor vest. Place the sensor vest next to the Bedside Console or in the
auxiliary basket if using a cart. This completes the measurement session.
Refer to the
Step 5, Reviewing Measurement Results
section on 57 for a description of the
results.
Unsuccessful Measurement
If the measurement was not successful, then the following message is displayed:
Low Quality
Verify that the instructions in this chapter were followed correctly. It is recommended to check
that the back sensor is pressed against the back, and that the patient does not move or talk
during the measurement process. For additional information you may refer to
Chapter 5,
Troubleshooting
on page 80.
Re-run the measurement process by taping on Redo button.