Setting Up ReDS System
21
2
Setting Up ReDS System
This chapter describes how to adjust the sensor vest to fit a specific patient and other general
setup procedures that must be performed one time prior to using the ReDS System.
CAUTIONS:
To ensure proper and dependable usage of the ReDS System, read all the instructions
provided in this user manual before using the ReDS System.
Do not operate the ReDS System without proper training. For additional information,
contact support.
General Setup
The following steps need only be performed once before using the ReDS System for the first
time.
Step 1: Assembling the Cart
This step is optional. If no cart is needed, skip to Step 2 on page 22.
To assemble the cart:
o
Assemble the cart stand wheels, basket and handle according to the assembly
instructions provided in the cart box, with the handle being positioned at a
distance of 1 cm from the top of the pole, and the basket positioned 78cm above
floor, both facing backwards (opposite to the VESA mounting plate).
o
Mount the Bedside Console on the VESA mounting plate, shown below. Use the
provided screws and a standard Phillips screwdriver.