Taking a Measurement
60
If this is the first time that the sensor vest is being used after it was adjusted, then repeated
unsuccessful measurements may indicate that the sensor vest was not fitted properly. In this
case, re-adjust the sensor vest by following the wizard’s instructions on the device. You can
also follow the instructions in
Chapter 2, Setting Up ReDS system
on page 21 and then re-run
the measurement process described in this chapter.
If the problem occurs three times in a row, remove the sensor vest and contact support.
Step 5: Reviewing Measurement Results
After a successful measurement process, the following is displayed on the Bedside Console
screen:
NOTE:
If the measurement was not successful, then the following message is displayed:
Low Quality
You may refer to page 59 for more information.
If the measurement was successful and if you are using the SensiCloud, the results are
transmitted at this stage. After a successful transmission, the following message is
displayed:
Your measurement was transmitted successfully
. In any case of transmission
failure, a message will be shown on screen saver and the results will be saved on the
device until transmitted successfully.
Regardless to transmission, the result will be saved on the device in Patient’s History
record.
For additional information regarding data transmission, you may refer to
Chapter 5,
Troubleshooting
on page 80.